Abstract

Fast low-angle shot (FLASH) imaging is a new technique for rapid magnetic resonance (MR) imaging that reduces acquisition times to seconds while retaining spatial resolution. This article deals with a three-dimensional (3D) variant of the FLASH method that allows the recording of a 3D-data set of 128 X 128 X 128 pixels within an acquisition time of only 4 min. The method is demonstrated using a 2.35 T 40 cm bore MR system. Experiments are carried out on rabbit head and human extremities. Depending on the field of view, the isotropic resolution is 1 mm or even less leading to cross-sectional images with a 1 mm slice thickness. In principle, FLASH imaging techniques are applicable to any MR system without the need of major hardware modifications. However, high-speed computers, large storage capacity, and rapid image display routines greatly facilitate an advantageous use of the 3D-FLASH variant.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.