Abstract
We report the design of an enhanced PeT-based fluorogenic phosgene probe 1-CN comprising a BODIPY fluorescence reporter, and a catechol quencher at the meso-position acting as the phosgene reaction site. The fluorescent catechol-derived probe 1-CN specifically reacts with phosgene to result in a cyclic carbonate ester. The fast (< 3 s) reaction suppresses a reductive PeT quenching process, leading to a fluorescence turn-on signal (F/F0 up to ca. 1300). The high sensitivity (LOD = 24 pM) of the probe is achieved by the precise adjustment of the energy levels of the frontier orbitals involved in the PeT.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.