Abstract
Four components of three sets of DNA constituents, bases, deoxyribonucleosides and deoxyribonucleoside 5′-monophosphate, were sufficiently resolved under one set of chromatographic conditions using high-performance liquid chromatography with a reversed-phase column (Zorbax ODS) and the solvent 0.4 M NH 4H 3PO 4 (pH 3.5). The effect of pH and salt concentration in the solvent on the retention of these compounds in the column was thoroughly investigated. Proportionality of peak height to the content, and reproducibility and recovery of the four bases were satisfactory under appropriate conditions and as little as 1 μg of DNA could be analysed for its base composition by this method.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Chromatography B: Biomedical Sciences and Applications
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.