Abstract

Direct-detected rapid-scan EPR signals were recorded using triangular field scan rates between 1.7 and 150 kG/s for deoxygenated samples of lithium phthalocyanine (LiPc) and Nycomed trityl-CD 3. These scan rates are rapid relative to the reciprocals of the electron spin relaxation times and cause characteristic oscillations in the signals. Fourier deconvolution with an analytical function permitted recovery of lineshapes that are in good agreement with experimental slow-scan spectra. Unlike slow-scan EPR, direct detection rapid-scan EPR does not use phase sensitive detection and records the absorption signal directly instead of the first derivative of the absorption signal. The amplitude of the signal decreases approximately linearly with applied magnetic field gradient. Images of phantoms constructed from samples of LiPc and trityl-CD 3 were reconstructed by filtered back-projection from data sets with a missing angle. The lineshapes in spectral slices from the image are in good agreement with slow-scan spectra and the spacing between sample tubes matches well with the known sample geometry.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.