Abstract

Chiral separations of 1-aminoindan (AI) by cyclodextrin electrokinetic chromatography (CDEKC) were investigated on microfluidic quartz chips. By using a microchip electrophoresis (MCE) instrument equipped with a linear-imaging UV detector, the separation process of the enantiomeric compounds was observed. When sulfated beta-cyclodextrin was employed as a chiral selector, the baseline separation of AI could be achieved within 1 min with a high repeatability. The relative standard deviation of the migration time was less than 6%. The fastest separation was achieved in 14 s, utilizing a separation length of only 6.1 mm. These results show that the MCE analysis employing a linear imaging UV detector has a significant potential for fast chiral analysis.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.