Abstract
A method for urinary iodine diagnostics using a portable energy-dispersion X-ray fluorescent (EDXRF) spectrometer is proposed. The principle of the method consists in excitation of the sample atoms fluorescence by the energy spectrum intentionally formed from the spectrum of an X-ray tube using an optimized X-ray scheme. The optimization by the criterion of the minimum detection limit for L-series iodine fluorescence lines included the calculation of optimal atomic numbers for materials of a re-radiator and a filter, and their thicknesses, the collimation system parameters, and an X-ray tube voltage. Experimentally achieved detection limits were 45 and 75 μg/L by I-Lα and I-Lβ lines, respectively, at the theoretically extreme value 30 μg/L. This sensitivity is found to be sufficient for urinary iodine diagnostics in the range from 50 to 200 μg/L. The results obtained from different patients have shown the satisfactory convergence for the iodine concentration determination by Lα and I-Lβ fluorescence lines. The simple sample preparation procedure and comparably small sizes of the apparatus allows rapid researches directly in clinical settings.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.