Stentless valves are designed to reduce mechanical stress and it is hypothesized that degeneration is reduced. We report early calcification of a Sorin pericarbon stentless pericardial xenograft in aortic position leading to valve failure 2 years after valve replacement in an 11-year-old boy. Morphological evaluation of the explant revealed severe calcification of the leaflets in a uniformly distributed pattern. Positive staining for non-collagen bone matrix proteins was found in the organic matrix of calcific deposits and in infiltrated macrophages. The stentless design of the Sorin valve does not mitigate calcification and therefore cannot be recommended for children.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.