Determination of intestinal permeability by measuring nonmetabolized sugars has been used to assess the integrity of intestinal mucosa. We have developed and validated a modified HPLC method for determining the concentration of L-rhamnose, mannitol, and lactulose in urine, using an amine-modified silica column and refractive index detection. Probe sugars are simultaneously resolved within 18 min. The calibration curve for each sugar is linear to 20 mmol/L. The minimum detectable concentration of lactulose is 0.05 mmol/L. Recovery of probe sugars is between 99.3% and 105.1%. Overall precision (CV) of estimation of probe sugars ranges from 4.2% to 6.5%. In 14 urine samples from healthy children who ingested the test solution containing 1 g of L-rhamnose, 1 g of mannitol, and 5 g of lactulose, the 5-h urinary excretion ratios of lactulose/rhamnose and lactulose/mannitol were 0.047 +/- 0.018 and 0.021 +/- 0.010 (mean +/- SD), respectively. This method presents a rapid, convenient, and practical technique for determining intestinal permeability in clinical pediatric practice.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.