Abstract
A sensitive and selective ultra performance liquid chromatography–mass spectrometric method has been developed for the quantitative analysis of nitrate in meat samples. Selected ion reaction (SIR) mode was adopted to identify and quantify the nitrate. Chromatographic analyses were performed on a BEH C-18 column with a mobile phase consisting of a surfactant (Cetylpyridinium chloride) and acetonitrile in equal ratio (50/50, v/v) at a flow rate of 0.4mLmin−1. The limit of detection and limit of quantitation of the developed method was found to be 0.0599 and 0.1817mgkg−1, respectively. The linearity of the proposed method was checked in the concentration range of 0.5–10mgkg−1 with an excellent correlation coefficient (r) of 0.997. The recovery of the nitrate in the meat samples were in the range of 98.02–98.99%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Spectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.