Abstract

Single crystals of synthetic orthopyroxene were analyzed by Raman spectroscopy and electron microprobe in order to calibrate the effect of chemical composition on the position of Raman bands. For the Raman peak close to 670 cm −1 we obtain the correlations between Raman shift (RS) and chemical composition FeO(wt%) = −1.998 · RS(cm −1 ) + 1373.2 and MgO(wt%) = 1.4537 · RS(cm −1 ) − 959.02, allowing the determination of FeO with an accuracy of ±0.4 wt% and MgO with an accuracy of ±0.3 wt%, corresponding to less than 1 mole% ferrosilite or enstatite. A test of this calibration using natural orthopyroxene crystals shows larger deviations up to 3 wt%. These deviations can mainly be related to significant amounts of CaO and Al 2 O 3 , leading to an overestimation of FeO or MgO, respectively.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.