Abstract

The radiometric efficiency of three analytical filters AFA-RSP-20, AFA-RMP-20 and AFA-RMV-20 was examined at different air velocities and aerosol number concentrations. An experimental setup had been constructed with 2 m3 radon chamber. Alpha radiometry was used to measure the deposited activities in the filters. Aerosols parameters like number concentration and size distribution are measured continuously with an aerosol diffusion spectrometer (ADS). RMV filter has a stable efficiency ~ 99.8%. The RSP and RMP filters efficiency is depending on the aerosol concentration and air sampling velocity. The effect of an aerosol concentration on the filter efficiency is more obvious than the effect of air velocity.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.