A numerical study to investigate the effects of the exposure to electromagnetic fields (EMF) at 900 and 1800 MHz on biological tissues implanted with thin metallic structures has been carried out, using the finite difference time domain (FDTD) solution technique. The results of the model show that the presence of a metallic wire yields to a significant increase in the local specific energy absorption rate (SAR). The present standards and/or guidelines on safe exposures of humans to EMF does not cover persons with implanted devices and thus the threshold levels to define safe exposure conditions might not apply in presence of high SAR gradients, such as the ones generated by thin metallic implanted objects. However, exposure to EMF fields below the actual safe levels even in presence of thin conductive structures cause rather low temperature rises (1 degrees C).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.