Abstract
In order to measure the radiation losses of a theta pinch plasma (electron densityne=1−5×1016 cm−3, electron temperatureTe=150−350 eV), a grazing incidence spectrograph is absolutely calibrated in the range 10–200 A. This is done in two steps: First the measured intensity ratios of lines emitted by hydrogen-like ions are compared with their calculated values thus yielding the relative sensitivity of the instrument. The result is confirmed by incorporating well known intensity ratios of lithium-like ions. Secondly absolute calibration is possible by hanging the spectrograph on an absolutely calibrated monochromator via the branching-ratios of lithium-like ions.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.