Abstract

In this paper, we report a novel quasi-common-path laser feedback interferometer (QLFI) for highly stable, high-resolution and non-contact displacement measurement. QLFI measures the displacement of the target by measuring the phase of feedback light. In addition to the target-generated feedback light (frequency shifted by 2#), a reference mirror generates a reference feedback light which is frequency shifted by #. The phase variations of both feedback lights are measured by heterodyne detection simultaneously and their difference offers the phase variations caused only by target displacement. When the optical path lengths of the reference and measuring feedback light are nearly the same, the phase fluctuations caused by the environment and laser instability are effectively removed. The heat-induced deformation of a He-Ne laser tube is successfully in-line measured using QLFI.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.