Abstract
A quantitative assay of IgA class circulating immune complexes (IgA-CIC) by a solid phase anti-C3 enzyme immunoassay (anti-C3 EIA) is described. A stable and reproducible standard for determination of IgA-CIC was prepared successfully by chemical binding of complement C3 to human serum IgA. Two of 27 sera from patients with systemic lupus erythematosus (SLE), however, contained high concentrations of IgA class anti-F(ab′) 2 antibodies that caused false positive results when the F(ab′) 2 of anti-C3 was used for EIA. Solid phase Facb of anti-C3 was found to eliminate the false positive results caused by IgA class anti-F(ab′) 2 and IgA class rheumatoid factor. Good reproducibility and recovery were observed with this Facb anti-C3 EIA using the IgA-C3, a stable standard material, and so this method should be useful clinically in elucidating the role of IgA-CIC.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.