Abstract
A simple, rapid, reliable and accurate HPLC method has been developed and validated as per the ICH guidelines for the quantitative determination of Donepezil HCl in bulk and tablets. The water HLC-10, high performance liquid chromatography system were used. The LOQ is detected at concentration 0.05mg/ml and the linearity range is 50-150g/ml. An optimum concentration scanned in the range 200-350 nm and ? max of the drug is 230 nm. No any interference at developed (R t = 8.22 min) and also the ? max peak is sharp. The Accuracy of the method was found to be in the range of 99.70% to 100.26%. The mean Inter and Intraday assay Relative Standard deviation (%RSD) were less than 0.69%. The Proposed method was found to be Linear, precise and accurate for the quantitative estimation of Donepezil in tablet formulations and can be used for commercial purposes.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.