Abstract

We examine the use of external standards for quantitative measurement by 1H NMR of solution concentrations of natural products and other low molecular weight, hydrogen-containing compounds and show that precision and accuracy ca. 1% is obtainable with a commercial 11.7 T spectrometer when standards and analytes are contained in separate but identical sealed precision glass NMR tubes. Numerous factors contributing to the intensity of the NMR signals are evaluated. Precise measurements of 360 degrees pulse lengths for each sample provide direct corrections for variations in probe Q-factor that enable samples in different solvents to be compared, provided single-coil excitation and detection is used throughout. Samples need not be prepared in deuterated solvents if the 1H spectra of the solvents are simple enough for peak suppression by presaturation. The approach is particularly suitable for hazardous materials kept in sealed tubes and for the preparation of certified calibration solution reference materials for use with LC-MS and other techniques where deuterated solvents should be avoided.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.