Abstract

Capillary zone electrophoresis is employed for the determination of caffeine using end-column amperometric detection with a carbon fiber microdisk array electrode at a constant potential of 1.45 V versus a saturated calomel electrode. The optimum conditions of separation and detection are 0.1 52mM NaH2PO4-0.648mM Na2HPO4 for the buffer solution, 20 kV for the separation voltage, 5 kV for the injection voltage, and 10s for the injection time. The limit of detection is 2.9 x 10(-4)mM or 1.2 fmol (signal-to-noise ratio = 2). The relative standard deviation is 0.68% for the migration time and 2.3% for the electrophoretic peak current. The method is applied to determining caffeine in human serum and a cola drink.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.