Abstract

Chemiluminescent analysers for measuring nitric oxide (NO) and nitrogen dioxide (NO2) in ambient air are generally calibrated with certified gas standard cylinders of NO in nitrogen. Verification of the NOx and NO amount fractions has been carried out on many such ‘on-site’ calibration cylinders at air quality monitoring stations. These measurements indicate that significant numbers of these gas mixtures have become somewhat degraded, with several percent of the NO oxidised to NO2. The effect of not compensating for this degradation on reported concentrations is discussed. If such degradation is not quantified and corrected for, there will be a systematic under-reporting of NO2 concentrations, which, due to the non-linearity of the effect, could reduce high reported NO2 concentrations at kerbside sites by around 20%. This could significantly reduce the number of reported exceedances of the NO2 limit value at such sites, compared to results obtained where there is no degradation of the NO cylinder.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.