Abstract
Hardware prominence remains a clinical challenge in focus for implant design in subcutaneous plate applications. Existing evaluation of hardware prominence relies on plate-to-bone distance at a single point or on average. A reproducible measure for plate prominence remains undefined. This study mathematically defines the plate prominence linked to the cross-sectional area change due to the plate presence on the bone. Two anatomical plate designs were fitted to 100 clavicles, and afterwards plate prominence parameters were evaluated and compared. This methodology enables the quantification of hardware prominence for different plate designs to inform the development of implants targeting low prominence.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Computer Methods in Biomechanics and Biomedical Engineering
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.