Abstract

The objective of this study was to evaluate the quality of five commercial enzyme linked immunosorbent assay (ELISA) kits (A, B, C, D, and E) from different suppliers for detecting aflatoxin B1 (AFB1). AFB1-free corn samples supplemented with different levels of AFB1 (5, 10, and 20 μg/kg) were used as positive controls and 6 replicates of each control sample were tested to evaluate the accuracy and precision of these kits. In addition, we also evaluated the performance of these ELISA kits for AFB1 in 30 feed samples, including corn, distillers dried grains with soluble, wheat samples, soybean meal, and poultry feed, which were verified by high performance liquid chromatography. Results showed that the coefficients of variation ranged from 1.18% to 16.22% in intra-plate and 2.85% to 18.04% in inter-plate for the determination of AFB1. The half maximal inhibitory concentration for five kits ranged from 3.72 to 7.22 μg/kg. The quantitation limits of AFB1 were all under the legal limit in China but somewhat inconsistent with kit instructions. Although the recovery rate of four of the five kits were either less than 90% or more than 110%, all these values were acceptable in practice. Two kits had high false positive rates (C and E). In conclusion, our results revealed that the qualities of five tested ELISA kits were significantly different.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.