Abstract

Three hundred and ten rapid-manufactured mechanical ventilators, named Masi, were produced and validated in Peru, according to applicable standards. From these, a sample of 30 was taken and two ventilation parameters, tidal volume and peak inspiratory pressure, were statically analyzed using control charts and histograms. Results show that several points were outside estimated limits for Shewhart means and ranges charts, which could possibly be due to the quantity of equipment used for data recollection and the fact that the Masi team had over 20 engineers. Nevertheless, Masi ventilators met the tolerance required by their user´s manual and MHRA standard and Peruvian DIGEMID for every parameter.Clinical Relevance-This article shows the performance in the validation stage of the peruvian mechanical ventilator MASI built as an emergency response for the COVID-19 crisis.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.