Abstract

Although semitendinosus-gracilis (SG) grafts for anterior cruciate ligament reconstruction have many potential benefits, effective fixation remains a challenge. This study assessed differences between the maximum pullout forces needed to detach a quadrupled SG graft from a femoral tunnel when secured by either a metal or a bioabsorbable interference screw. Sixteen paired fresh frozen quadrupled SG autografts (x, 8.4 mm; range, 7-10 mm) from 8 donors were anchored into cadaveric femoral tunnels by either a 7-mm metal or a 7-mm bioabsorbable screw after bone mineral density (BMD) assessment and before undergoing a longitudinal 20 mm/min traction force. Statistical analysis compared BMD (gm/cm2), insertional torque (N-m), and maximum load at pullout (N) between screw types. Insertional torques (.28-1.21, N-m range) did not correlate ( P>.05) to BMD or maximum load at pullout (x ± S.D.) 242 ± 90.7 N (metal screw) and 341.1 ± 162.9 N (bioabsorbable screw). Differences did not exist between the maximum load at pullout for bioabsorbable or metal screw fixation ( P = .16). Careful graft preparation, sizing, and matched tunnel placement enables interference fit and fixation capable of reliably withstanding the low- level rehabilitation loads to which the graft is exposed until bony ingrowth occurs. Arthroscopy 1998 Apr;14(3):241-5

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.