Abstract

The article provides an overview of the measurement issues affecting sensing systems based on quartz crystal microbalances (QCM), operating in liquids. The article analyzes the effect of the presence of a viscous liquid at the interface with the QCM sensors by analytical models and experimental tests based on impedance measurements in reference conditions. The study takes also into account the parasitic effects due to the measurement setup and provides results, which can be applied to a wide range of scenarios also in the presence of highly viscous media. The article also presents a model-fitting procedure for QCM impedance spectra, which provides accurate estimations of the electromechanical parameters.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.