Abstract

Multichannel patient monitors apparently display the electrocardiogram, intra-arterial blood pressure, and pulse oximetry waveforms in synchrony during sinus cardiac rhythm. We recorded 20 events of a premature cardiac contraction on multichannel monitors from 10 patients. Interestingly, during premature cardiac contraction, the low-amplitude arterial trace was aligned with the premature electrocardiogram, but the low-amplitude pulse oximetry plethysmograph was aligned with the next normal electrocardiogram and arterial tracing. In conclusion, the pulse oximetry plethysmograph tracing is offset by 1 ventricular depolarization on the electrocardiogram (QRS) and the arterial tracing on multichannel monitors.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.