Abstract

Pulse inversion harmonic imaging (PIHI) is a new modality that increases the detection of harmonic echoes and myocardial contrast by canceling linearly transmitted signals. We tested whether PIHI improved the detection of endocardial borders in noncontrast 2-dimensional echocardiography. We compared PIHI with tissue harmonic imaging (THI), which decreases linearly transmitted signals using filters. Fundamental mode (FM) was compared with THI and PIHI in 50 consecutive patients. The global and segmental endocardial visualization scores measured with FM were significantly improved by using either THI or PIHI. The improvement of the global score compared with FM was slightly higher using PIHI than THI, because of an improved visualization of the base and the anterior wall with the PIHI technique compared with THI. The ratio of myocardial-to-cavity signal was similarly increased from FM with THI and PIHI. PIHI, a new modality for detection of myocardial contrast, can also be used for endocardial border visualization. It provides an improvement relative to THI for specific regions of the endocardium. (J Am Soc Echocardiogr 2002;15:302-8.)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.