Abstract

Pseudo-continuous arterial spin labeling (ASL) can provide best signal-to-noise ratio efficiency with a sufficiently long tag at high fields such as 7 T, but it is very sensitive to off-resonance fields at the tagging location. Here, a robust Prescan procedure is demonstrated to estimate the pseudo-continuous ASL radiofrequency phase and gradients parameters required to compensate the off-resonance effects at each vessel location. The Prescan is completed in 1-2 min and is based on acquisition of label/control pair-wise ASL data as a function of the radiofrequency phase increment applied to the pseudo-continuous ASL train. It is shown that this approach can be used to acquire high quality whole-brain pseudo-continuous ASL perfusion data of the human brain at 7 T.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.