Abstract
Slides for ultra thin-layer chromatography (UTLC) were made by coating nonporous silica particles, chemically modified with polyacrylamide, as 15 μm films on glass or silicon. Three proteins, myoglobin, cytochrome c and lysozyme, are nearly baseline resolved by the mechanism of hydrophilic interaction chromatography. A plate height as low as 3 μm, with 3900 plates, is observed in 14 mm. Varying silica particle diameter among 900, 700 and 350 nm showed that decreasing particle diameter slightly improves resolution but slows the separation. Matrix-assisted laser desorption/ionization (MALDI)-MS of the proteins after separation is demonstrated by wicking sufficient sinapinic acid into the separation medium.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.