Abstract

Left ventricular assist device (LVAD) implantation has become an effective treatment for end-stage heart failure patients. However, LVAD support disrupts the natural intracavitary blood flow path through the heart, introducing flow patterns potentially associated with thrombosis especially around the inflow cannula. We present initial insights into in vitro 4D flow magnetic resonance imaging (MRI)-based analysis of flow patterns and an enzyme-activated milk test in a porcine left ventricular (LV) model during LVAD support, in a standard implant technique and in combination with a novel cone-shaped LV sparing prosthesis. 4D MRI of an LVAD-supported ex vivo porcine heart failure model was performed in a mock circulatory loop with an LVAD flow of 4.5 l/min. Two separate sets of hearts with an MR-compatible LVAD inflow cannula were used: one with a standard inflow via the pulmonary veins into the LV, one sparing the LV with the cone-shaped prosthesis sutured to the mitral annulus and the LVAD inflow cannula sparing the LV cavity. In correlation, an enzyme-activated milk test was performed to reveal locations of disturbed flows. Data were analysed visually and by performing a case-by-case analysis of flow patterns and enzyme activation locations. Using 4D flow MRI, the standard implant technique showed markedly altered flow patterns with flow velocity reduction in the LV cavity especially around the inflow conduit. The average flow velocity in the LV cavity was 12.8 ± 2.9 m/s. In contrast, the LV-sparing technique with a cone-shaped prosthesis showed faster and more organized flow of 36.3 ± 7.2 m/s on average to the LVAD inflow cannula. Compared with the standard technique, no recirculating blood flow components were detected. The enzyme-activated milk test confirmed the location and extent of helical and vertical flow patterns. For the first time, intracardiac flows in an LVAD setting have been shown in vitro based on a mock porcine ex vivo heart failure model by applying 4D flow MRI. Differences in flow characteristics of the same model using a prosthesis sparing the LV cavity are encouraging to further investigate their potential to reduce potential LVAD-associated side-effects, such as intracavitary clotting and its sequelae.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.