In this work, the performance of direct-load-measuring transducers of the recent “pressure-sensitivepaint” variety is critically analyzed. The characteristics of laboratory-prepared and commercial load cells incorporating this paint are studied. For the transducers tested, the conclusion reached was that these cells are unreliable in their load-indicating properties. The application of pressure-sensitive paints of this type in the measurement of other quantities such as pressure, strain and acceleration would appear unjustified because of the poor results obtained in the measurement of simple, direct loads.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.