Abstract
ABSTRACTThe pressure effect on the electromotive force (EMF) of a Pt13Rh–Pt (type R) thermocouple was examined to determine the temperature measurement accuracy of solid pressure medium apparatuses in high-pressure experiments. Single-wire EMFs were measured up to pressure of 13 GPa and temperature of 1173 K with a Kawai-type multi-anvil apparatus for Pt13Rh and Pt based on the single-wire method. The pressure conditions along the wires were evaluated by in situ X-ray diffraction using synchrotron X-ray radiation. The pressure effect of the Seebeck coefficients of Pt13Rh and Pt were determined by the analysis of the single-wire EMFs and pressure–temperature profiles along the wires and was virtually consistent with those determined in previous studies at lower pressures and temperatures. For type R thermocouple, the difference between the nominal and real temperatures was determined to be as large as –75 K at 13 GPa and 873 K.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.