In this research, we successfully designed a capacitor plate as a Plethysmometer sensor with a measured volume capacity of 1.5 ml, this capacity is obtained from a capacitor plate with a thickness of 0.5 mm width of 15 mm and a height of 200 mm. The sensitivity obtained is 0.36 / 0.1ml with a standard deviation of 0.012. The above results show that the linear capacitor plate sensor for the tested area has sufficient sensitivity that can measure the change of 0.1 ml to be used as a transducer in the Plethysmometer.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.