Abstract

Initial proof-of-concept development of competitive lateral flow (LF) test strips involving red quantum dots (Qdots) is demonstrated with visible confirmation of the competitive displacement of Qdot-protein-saxitoxin (Stx) and -tetrodotoxin (Ttx) conjugates over 15min. The measured limits of detection (LODs) for the best versions of the assays were in the 1-2µg range when a bovine serum albumin (BSA)-Stx or -Ttx conjugate was used as the analyte for safety reasons and NIH ImageJ image analysis was applied. However, when one assumes that only one of the primary amines in BSA labeled with Stx or Ttx is needed as the "epitope" for capture and calculates the weight of the toxin adducts on an equal mole basis to that of 1µg of toxin-BSA conjugate, the LODs are as low as 4.5ng of toxin. No cross-reactivity was seen for these prototype test strips when BSA-Stx and BSA-Ttx were switched as analytes at or near their respective LODs which is in agreement with published specificity and affinity studies for these aptamers in the literature.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.