Abstract
In vivo biodistribution imaging of platinum-based compounds may allow better patient selection for treatment with chemo(radio)therapy. Radiolabeling with Platinum-195m (195mPt) allows SPECT imaging, without altering the chemical structure or biological activity of the compound. We have assessed the feasibility of 195mPt SPECT imaging in mice, with the aim to determine the image quality and accuracy of quantification for current preclinical imaging equipment. Enriched (>96%) 194Pt was irradiated in the High Flux Reactor (HFR) in Petten, The Netherlands (NRG). A 0.05M HCl 195mPt-solution with a specific activity of 33MBq/mg was obtained. Image quality was assessed for the NanoSPECT/CT (Bioscan Inc., Washington DC, USA) and U-SPECT+/CT (MILabs BV, Utrecht, the Netherlands) scanners. A radioactivity-filled rod phantom (rod diameter 0.85-1.7mm) filled with 1MBq 195mPt was scanned with different acquisition durations (10-120min). Four healthy mice were injected intravenously with 3-4MBq 195mPt. Mouse images were acquired with the NanoSPECT for 120min at 0, 2, 4, or 24h after injection. Organs were delineated to quantify 195mPt concentrations. Immediately after scanning, the mice were sacrificed, and the platinum concentration was determined in organs using a gamma counter and graphite furnace - atomic absorption spectroscopy (GF-AAS) as reference standards. A 30-min acquisition of the phantom provided visually adequate image quality for both scanners. The smallest visible rods were 0.95mm in diameter on the NanoSPECT and 0.85mm in diameter on the U-SPECT+. The image quality in mice was visually adequate. Uptake was seen in the kidneys with excretion to the bladder, and in the liver, blood, and intestine. No uptake was seen in the brain. The Spearman correlation between SPECT and gamma counter was 0.92, between SPECT and GF-AAS it was 0.84, and between GF-AAS and gamma counter it was0.97 (all p < 0.0001). Preclinical 195mPt SPECT is feasible with acceptable tracer doses and acquisition times, and provides good image quality and accurate signal quantification.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: European journal of nuclear medicine and molecular imaging
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.