We present a new three-step auto-alignment algorithm for the specimen stage of an ellipsometer with adjustable angle of incidence. Correction of errors in tilt angle and position of the specimen stage can be performed by locating the reflected light spot at the center of the detector at two different angles of incidence. The current method does not need auxiliary focusing equipment. The alignment algorithm works to high precision in both model simulation and practical experiments with a rotating analyzer ellipsometer.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.