Abstract
Although image-guided radiotherapy (IGRT) is widely used to determine and correct daily setup errors, the additional interpretation for image registration would provide another error. We evaluated the uncertainty in image registration in IGRT. The subjects consisted of 12 consecutive patients treated with IGRT for thoracic esophageal cancer. Two radiation therapists had consensually achieved daily 3D registration between planning computed tomography (CT) and cone beam CT (CBCT). The original data sets of image registration in all fractions except for boost irradiations with a change in the isocenter positions were selected for evaluation. There were 20 to 32 data sets for each patient: a total of 318 data sets. To evaluate daily setup errors, the mean 3D displacement vector was calculated for each patient. To assess the reproducibility of image registration, two other radiation therapists reviewed the data sets and recorded geometric differences as uncertainty in the image registration. The mean 3D displacement vector for each patient ranged from 4.9 to 15.5 mm for setup errors and 0.7 to 2.2 mm for uncertainty in image registration. There was a positive correlation between the 3D vectors for setup error and uncertainty in image registration (r = 0.487, p = 0.016). Although IGRT can correct the setup errors, potential uncertainty exists in image registration. The setup error would disturb the image registration in IGRT.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.