Abstract

To illustrate the potential for high permittivity materials to be used in decreasing peak local SAR associated with implants when the imaging region is far from the implant. We performed numerical simulations of a human subject with a pacemaker in a body-sized birdcage coil driven at 128 MHz with and without a thin (5 mm) shell of material of high electric permittivity around the head. For a shell with relative permittivity of 600, the maximum specific energy absorption rate averaged over any 1 g of tissue near the pacemaker was reduced by 73.5% for a given B1 field strength at the center of the brain. Although further study is required, initial simulations indicate that strategic use of high permittivity materials may broaden the conditions under which patients with certain implants can be imaged safely. Magn Reson Med 78:383-386, 2017. © 2016 International Society for Magnetic Resonance in Medicine.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.