Abstract
The purposes of this study were to evaluate the low-contrast detectability of CT images assuming hepatocellular carcinoma and to determine whether dose reduction in abdominal plain CT imaging is possible. A Catphan 600 was imaged at 350, 250, 150, and 50 mA using an Aquilion ONE PRISM Edition (Canon) and reconstructed using deep learning reconstruction (DLR) and model-based iterative reconstruction (MBIR). A low-contrast object-specific contrast-to-noise ratio (CNRLO) was measured and compared in a 5-mm module with a CT value difference of 10 HU, assuming hepatocellular carcinoma; a visual examination was also performed. Moreover, an NPS within a uniform module was measured. CNRLO was higher for DLR at all doses (1.12 at 150 mA for DLR and 1.07 at 250 mA for MBIR). On visual evaluation, DLR could detect up to 150 mA and MBIR up to 250 mA. The NPS was lower for DLR at 0.1 cycles/mm at 150 mA. The low-contrast detection performance was better with DLR than with MBIR, indicating the possibility of dose reduction.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.