Abstract

The mechanical property of the BaSrFBr:Eu phosphor layer of X-ray image plates was investigated by using resolution (LP/mm) and coincidence Doppler broadening (CDB) positron annihilation as well as positron annihilation lifetime (PAL). The image plate samples of BaSrFBr:Eu phosphors in this experiment were irradiated by using hospital X-rays. The LP/mm values of the irradiated BaSrFBr:Eu image plates varied from 3.35 to 1.25 for up to 20,000 exposures. CDB positron annihilation and lifetime spectroscopy were used to analyze defect structures in the phosphor layer. Even when the LP/mm values were greatly decreased due to exposures, the S parameter and the lifetime (τ1, τ2) values were almost constant with increasing number of exposures. A positive relationship existed between the SEM images and positron annihilation spectroscopy (PAS). According to the SEM images and the positron annihilation spectroscopy (PAS) results, measurements of the defects with PAS indicate that the image-plate phosphor can be safely used for hospital X-rays in the course of diagnostic radiography at an average rate of 20,000 times for one year.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.