Abstract
We describe a novel position mapping and the uniformity correction method to improve the performance of small animal single-photon emission computed tomography (SPECT) imaging. The SPECT system consists of a cerium doped lutetium−yttrium oxyorthosilicate (LYSO) scintillation crystal (22×22 pixel array, 2mm×2mm×3mm pixel size), a position sensitive photomultiplier tube (H8500c, Hamamatsu Photonics Co., Ltd., Shizuoka Prefecture, Japan), and a parallel-hole collimator (Nuclear Fields Pty. Ltd., St. Marys, Australia). The position mapping method was based on a connected regional recognition algorithm. We present the algorithm and step-by-step details of image boundary detection, dynamic binarization, connected regional recognition, center-of-gravity computing, and look-up table establishment. The position mapping and uniformity correction tables were generated and applied to the SPECT projection data. The corrected projection images demonstrated that this correction method improved the uniformity of the raw projection image by ∼16%. The preliminary SPECT reconstruction results (using algebraic reconstruction technology) are also presented. A comparison between the reconstructed images before and after correction further confirms the performance of this correction method.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.