Abstract

All pharmaceutical manufacturers are required to verify that their production equipment is free from contaminants. Here, we report the capability of a fully automated portable capillary electrophoresis instrument with an integrated sample swab extraction – the Grey Scan ETD-100 – for the detection of pharmaceutical residues on surfaces of manufacturing equipment. Lidocaine was used as a model compound and could be recovered from a surface by swabbing, extracted from the swab, and analysed within 1 min. The recovery of lidocaine from a stainless-steel coupon was 81.3 %, with a LOD of 0.13 µg/swab. This fast, sensitive, and simple method implemented on a user-friendly portable CE instrument without the need for manual sample pre-treatment provides the possibility for on-site rapid determination of equipment cleanliness in the pharmaceutical industry.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.