Abstract
We were able to simplify the structure of the autocollimators and enhance the performance by using one- and two-axis PSDs (position-sensitive detectors) which have spatial resolution of 0.1 and 2.5 μm, respectively, with little dependence on the spot size of incident beam and the visible laser diodes as the light source. Two kinds of portable autocollimators which can measure angles with resolutions 0.01″ and 0.1″, respectively, were made. They were calibrated by the reference angle generator which employed the laser interferometer. The characteristics of temporal stability were evaluated and the profile of a test surface was measured.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.