Abstract

To evaluate the feasibility and safety of rotablation atherectomy in a suboptimally expanded stent. Seven pigs underwent suboptimal stent expansion in the left anterior descending coronary. Pulverization of the stent struts was performed by rotablation atherectomy with two different burr sizes. Two types of control porcine models were used: pigs with fully expanded stents and pigs without stents. Continuous electrocardiogram readings as well as microscopic and radiologic analysis of cardiac tissue were performed. Rotablation atherectomy reduces the suboptimally expanded stent by (26.95 ± 5.03)%. Ninety-five percent of the metal microparticles, imbedded in the suboptimally expanded stent group cardiac tissue, are less than 15 μm. Transient vasospasm and ST segments elevations were observed during rotablation atherectomy, which returned to basal conditions at the end of the intervention. Our study demonstrated the feasibility and safety of using rotablation atherectomy to pulverize stent struts in a suboptimally expanded stent.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.