The specific feature of NQR is expansion of spectral lines which is caused not only by dipole-dipole interaction of nuclei, but also by local field nonuniformity caused by the defects and deformation in crystal matrix. Considerable line expansion, which is typical of crystals, requires in pulsed NQR method the optimization of pulse shape and the reserve of transmitter power output. Parametric computer identification was used to study a dependence of parameters of the energy spectra of the output signal of pulsed NQR spectrometer transmitter on the duration of excitation pulses with sine and noise occupation. The energy efficiency of a linear amplifier was calculated and experimental investigations of its temperature conditions were carried out. The energy-efficient broadband transmitter was proposed that can be used in portable setups for the pursuance of research in the field of pulsed NQR spectroscopy, for instance when studying isotopes with quadrupole moments 14N, 35Cl, 63Cu, 69Ga, 71Ga, 113In, 115In and others.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.