Abstract

The evaluation of the size-exclusion chromatography (SEC) concentration elution curves by means of a calibration dependence obtained in a given SEC set for a polymer different from the polymer to be analyzed results in an error in the determination of both molecular weight and molecular-weight distribution (MWD). The problem is analyzed assuming the validity of the universal-calibration concept. The differences between the true and apparent values of molecular weight, MWD and M w/ M n depend on and are expressed in terms of the parameters of the Mark–Houwink–Kuhn–Sakurada equation, describing the molecular-weight dependence of intrinsic viscosity, for the polymer to be analyzed and the polymer used for calibration. The differences in molecular weight and the M w/ M n ratio are typically tens of percent and, in extreme cases, can amount up to a factor of three for molecular weight and a factor of two for the M w/ M n ratio.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.