Abstract

We developed polarization fluoroimmunoassays for 11-deoxycortisol in serum and saliva. To avoid interfering factors, the steroid was initially extracted from the biological fluids with dichloromethane. Assays could then be completed without any further separation procedures or need to correct for blank signals. The serum assay was suitable for following the response to the metyrapone test and results correlated acceptably with those by an established, specific, direct 125I-radioimmunoassay. The method is not sufficiently sensitive to detect 11-deoxycortisol in normal saliva, but greatly increased concentrations were found in post-metyrapone saliva and results agreed well with those by the radioimmunoassay as modified for salivary assay. Salivary 11-deoxycortisol assay would provide a convenient means of monitoring results of the metyrapone test.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.