Abstract

Plasma-spraying of metallic impiant surfaces is an established method for the application of hydroxylapatite (HA) coatings. Carbon fibre reinforced thermoplastics show different thermal and mechanical properties, compared with titanium substrates. In this paper first results of the influence of the established coating method on carbon fibre reinforced thermoplastics are presented. First investigations of the tensile adhesion strength, tested with a newly developed testing device, showed that the adhesion between the HA coating and the carbon fibre reinforced polyetheretherketone (PEEK) composite is very low. Macromechanical bending tests showed a change to initial tensile instead of compression failure of the coated composite substrate. Micromechanical bending tests in a scanning electron microseope (SEM) hot tensile stage (Raith GmbH) revealed crack propagation within the ceramic coating and in the coating-substrate interface before the total failure of the composite substrate occurred.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.