Abstract

Background: Test chambers for irritant patch testing are usually larger than those used in allergic patch testing. In general, larger areas show stronger skin reactions than smaller areas. Objective: This study investigated whether this difference is of practical relevance, when a model irritant is applied in small and large Finn chambers and evaluated by measurement of transepidermal water loss (TEWL). Methods: Patch testing was performed with 2 concentrations of sodium lauryl sulfate (SLS) (0.25% and 0.5%) on forearms of healthy volunteers. Large (inner diameter, 12 mm) and small (inner diameter, 8 mm) chambers were used. Results: A variance analysis (3 factors, 2-tailed) showed that the test outcome, as assessed by TEWL, was strongly dependent on SLS test concentration and test chamber size. The larger chambers gave approximately 30% to 50% higher values than the smaller. Conclusions: This may be explained by the fact that with the small chambers, the adjacent small area of nontreated skin was also assessed by the evaporimeter, biasing the results. A formula estimating TEWL value of the large chamber from values of the small chambers has been proposed.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.