Abstract

Central-axis percentage depth doses and tissue-maximum ratios (TMR) for 45-MV photon beams from a betatron have been measured in water. Also the influence of field size and collimator scatter on the dose in the buildup region have been investigated. The maximum dose for TMR has been shown to occur at a point about 2 cm deeper than the maximum dose for percentage depth dose. This difference is significant in the characterization of the photon beam at this high energy. The measured physical data have been computerized for use in routine treatment planning. Computer-gererated beams have been found to be in close agreement with measured isodose curves. Computer-generated isodose distributions for typical clinical irradiation techniques have been verified using RP/V film in an Alderson phantom.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.