Abstract

In hadron therapy one of the most advanced methods for beam delivery is the active scanning technique which uses fast scanning magnets to drive a narrow particle beam across the target. The Centro Nazionale di Adroterapia Oncologica (CNAO) will treat tumours with this technique. The CNAO scanning system includes two identical dipole magnets for horizontal and vertical beam deflection, each one connected to a fast power supply. The dose delivery system exploits a set of monitor chambers to measure the fluence and position of the beam and drives the beam during the treatment by controlling the sequence of currents set by the power supplies. A test of the dynamic performance of the scanning system has been performed using a Hall probe to measure the field inside the magnet and the results are presented in this paper.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.